Technical data
- Process temperature
-
- Refer to section “Operating conditions”
- Pt100 accuracy class (EN 60751)
-
- 1/1 B ± (0.3 + 0.005 × t)°C
- 1/1 A ± (0,15 + 0,002 × t)°C
- 1/3 B ± 1/3 × (0.3 + 0.005 × t)°C
- 1/6 B ± 1/6 × (0.3 + 0.005 × t)°C
- Thermal response time, T90
-
- RTD element only
- ≤ 3.0 s , Ø3 mm
- ≤ 3.6 s , Ø4 mm
- ≤ 8.5 s , Ø6 mm
- Thermal response time, T50
-
- RTD element only
- ≤ 1.0 s , conical sensor
- ≤ 1.3 s , Ø3 mm
- ≤ 2.5 s , Ø4 mm
- ≤ 3.0 s , Ø6 mm
- Process pressure
-
- Refer to section “Operating conditions”
- Sensor length
-
- Sensor tube material
-
- Surface roughness wetted parts
-
- Degree of protection (EN 60529)
-
- IP 65 , with DIN EN 175301-803 A (DIN 43650 A), 4-pin
- IP 68 , with M12-A, 4-pin and appropriate cable
- Humidity
-
- Insulation resistance
-
- > 100 MΩ , 500 V DC (for Rail version)
- Insulation voltage
-
- 600 V AC , EN 50155 (for Rail version)
- Long-term test (transmitter)
-
- Vibration (sinusoidal) (EN 60068-2-6)
-
- 1.6 mm p-p (2 … 25 Hz), 4 g (25 … 100 Hz), 1 octave / min.
- Without transmitter
-
- With transmitter
-
- EMC
-
- EN 61000-6-2
- EN 61000-6-3
- EN 61326-1
- NE21
- Hygiene
-
- FDA (21 CFR 177.2415)
- 3-A (74-07)
- Railway applications
-
- Explosion protection
-
- Ex ia Simple apparatus, gas and dust